Top-grade performance with the high resolution 640 x 480-pixel detector!
Type | H2640 | H2630 (discontinued) |
---|---|---|
Measuring range |
-40 to 500°C Range1: -20 to 60°C Range2: -40 to 120°C Range3: 0 to 500°C Range4: 200 to 2000°C optional (Display 0 to 2000°C) |
-40 to 500°C Range1: -40 to 120°C Range2: 0 to 500°C Range3: 200 to 2000°C optional (Display 0 to 2000°C) |
Resolution |
Range1: 0.06°C or better (at 30°C, 30Hz) 0.03°C or better (at 30°C, Σ64) Range2: 0.08°C or better (at 30°C, 30Hz) Range3: 0.12°C or better (at 30°C, 30Hz) |
Range1: 0.08°C or better (at 30°C, 30Hz) 0.04°C or better (at 30°C, Σ16) Range2: 0.12°C or better (at 30°C, 30Hz) |
Accuracy | ±2°C or ±2% of reading, whichever is greater | |
Detector | Uncooled focal plane array (microbolometer) | |
Spectral range | 8 to 13μm | |
I.F.O.V | 0.6mrad | |
Focusing range | 30cm to infinity | |
Field of view | 21.7°(H)x16.4°(V) | |
Frame time | 30 frames/sec * | |
Display | View finder & 5.6-inch movable color LCD | |
Thermal image pixels | 640(H)x480(V) pixels | |
A/D resolution | 14bits | |
Focus system | Auto and manual focus (mechanical/motorized) | |
Alarm | Screen display and alarm sound (ON/OFF) | |
S/N improvement | Σ2, Σ8, Σ16, Σ32, Σ64 | Σ2, Σ8, Σ16 |
Real-time memory (H2640 only) | Multi-speed to max. 30fps* continuous shooting up to 832 images | |
Interval measurement |
Recording on built in real time memory: 1/30 to 3600 sec interval (H2640) * Recording on memory card: 5 to 3600 sec interval (thermal image) 30 to 3600 sec (thermal&visual images) Trigger function provided |
|
Emissivity correction | Provided (0.10 to 1.00). Emissivity table provided. | |
Env. temp. correction | Provided (including interval NUC) | |
User setup | Pre-registration of environmental setup (up to 10 setups) | |
Background compensation | Provided | |
Distance correction | Provided | |
Auto functions | Full automatic (level, sense, focus), Level trance, Auto-gain control | |
Thermal / Visual image fusion | Provided | |
Visible light camera | 1.31 Mega pixels, focusing distance: 30cm to infinity | |
Display functions |
Display color: color / monochrome, positive / negative Gradation: 16, 32, 64, 128, 256 Color palette: rainbow, brightness, shine, hot-iron, medical, fine Isothermal band display: max. 4 bands, movable Thumbnail display: 12 thermal images replay Image rotation: 180° rotation Multi-sense display, Battery life indicator Waveform display: X-Y line waveform at Freeze mode Multilingual menu |
|
Digital zoom | x2, x4, x8 | x2, x4 |
Multi-point temperature display | Up to 10 points | Up to 5 points |
Multi-point emissivity correction | Up to 10 points | Up to 5 points |
Multi-focus ** (H2640 only) |
On/Off, temp threshold (1 to 100°C, 1°C step), Images to capture (16 or 32), direction of edge detection (horizontal, vertical, horizontal & vertical) |
|
Image processing functions |
Variable level / sense Temperature difference between 2 points (Δt) Max / Min (peak hold) temperature display Alarm (full screen or specified box) BOX setting, up to 5 boxes (max, min, average) Sharpness filter, Median filter, Average filter |
|
Annotation | Text and voice annotation (30 sec per image) | |
Storage device |
Compact flash memory card for: Thermal image in SIX or BMP file format visual image in SIX or JPEG file format |
|
video signal output | NTSC / PAL composite, S-video (optional cable TH91-348 required) | |
Interfaces |
IEEE1394a, RS-232C, USB1.1 / 2.0 (Mass-storage) Compact flash memory card Slot |
|
Operating temperature | -15 to 50°C, 90%RH or less (not condensed) | |
Storage temperature | -40 to 70°C, 90%RH or less (not condensed) | |
Power supply | AC adaptor: DC 7.2V (nominal) | |
Power consumption | Approx. 9W (Typ) | |
Shock & vibration |
Shock: 294m/s2 (30G) (IEC60068-2-27) Vibration: 29.4m/s2 (3G) (IEC60068-2-6) |
|
Water/Dust proof rating | IP54 (IEC60529) | |
Dimensions & weight |
Approx. 110(W)x110(H)x210(D)mm (excluding projections) Approx. 1.5kg (without battery, LCD) Approx. 1.7kg (with battery, LCD) |
|
Software included | Thermal image capture program & NS9500LT | NS9500LT Report generator (Limited version) |
Standard accessories: AC adapter, battery pack(2pcs), battery charger, compact flash memory card & adapter, USB cable, target illuminator, carrying case, grip belt, neck strap, lens cap, operation manual
Model | Product Name | Description |
---|---|---|
NS9200 | Report generator Software | Requires Windows XP(professional) Service Pack 2(or newer) or Windows 2000(professional) Service Pack 4(or newer). Microsoft Excel and Word (2000, 2002 or 2003) are also required for operation of some functions* |
TH92-717 | IEEE1394 data capture software with cable | For real-time image data transfer via IEEE1394, comes with 4m cable |
Model | Product Name | Description |
---|---|---|
TH71-334,359,360 | AC adapter | Safety standard: PSE, UL, CSA (334: 110V PSE, 359: 220-240V, 360: 110V UL) |
TH91-465 | Battery pack | Lithium ion battery |
TH71-339,340 | Battery charger | With 2 battery slots (339: 200-240V, 340: 100-110V) |
TH92-490 | High temperature range | 200 to 2000°C |
TH92-481 | 5°telephoto lens (4.5x) | 4.9°x3.7° focusing range: 20m to infinity |
TH92-482 | 10° telephoto lens (2x) | 10.9°x8.2°with visual camera, focusing range: 2m to infinity |
TH92-483 | 42°wide-angle lens (0.5x) | 45.2°x33.7°with visual camera, focusing range: 0.3m to infinity |
TH92-486 | 100μm close-up lens | Min. detectable size: 100x100μm, scan range: 64x48mm, working distance: 195mm (Add-on type) |
TH92-485 | 25μm close-up lens | Min. detectable size: 25x25μm, scan range: 16x12mm, working distance: 11mm (Add-on type) |
TH92-488 | Lens adapter | For attaching TH9100 close-up lenses to H2600 (Applicable lenses: TH91-385 / 386) |
TH92-491 | 2nd year product warranty | Except temperature calibration |